Intuitive user navigation, preset therapy programs and overall flexibility. SIROLaser Xtend gets right to work. I bet that soon you couldn't imagine life without lasers in your practice!
Configuration
-Preset therapy programs, help menu, 24 freely programmable applications
-Access codes for 2 users
-1 handpiece
-Available as an upgrade: Hotkeys
-for 12 favorites, treatment documentation, rechargeable battery for mobile use
-Power: up to 10 watts
Technical data:
-Beam guide: Flexible quartz glass fiber
-Display: Full color, graphical LCD touch screen
-Cooling: Inteal air cooling controlled by output
-Temperature switch: Software temperature switch at 48° C
-Door contact connection: Potential-free contact 5 VDC/20 mA (TTL)
-Dimensions (W x L x H): 182 x 197 x 189 mm
-Weight: approx. 1000 g incl. handpiece
SIROLaser Xtend specifications:
-Laser type: Diode laser
-Wavelength: 970 nm ± 15 nm
-Laser system: Class IV (according to IEC 60825-1)
-Device classification: Class IIb (according to Council Directive 93/42/EEC)
-Optical power (CW): SIROLaser Xtend: approx. 0.5 - 5.0 W
After upgrade: approx. 0.5 - 7.0 W
-Optical peak power (diode): SIROLaser Xtend: approx. 10 W
After upgrade: approx. 14 W
-Emission modes: CW (continuous wave)
chopped SIROLaser Xtend: 1 Hz to 1.000 Hz
chopped after upgrade: 1 Hz to 10 kHz
peak-pulse approx. 1.5 kHz - 20 kHz
-Pulse: repeated pulse
-Pulse duration: Chopped mode SIROLaser Xtend: 500 µs - 0.5 sec.
Chopped mode after upgrade: 10 µs - 0.99 sec.
Peak-pulse: 23 µs fixed
-IP degree of protection: Laser unit: IP20
-Aiming beam: 635 - 650 nm, max. 1 mW
-NOHD: From the distal end of the optical fiber: 1.5 m
-Optical fiber thickness: 200 and 320 µm
-Operation: Finger switch, with electronic access key
-Nominal power input: 15 V DC 6.66 A max. 100 VA MPU100-106
-Insulation class: Class I, type B (according to IEC 60601-1)
-Adapter: Exteal, 100 - 240 VAC, 50 - 60 Hz
-Power supply: The SIROLaser Xtend may only be operated with the Sinpro MPU100-106 power supply.
Included:
-1x SIROLaser Xtend control unit including handpiece with integrated finger switch
-1x Starter kit fibers, 1 x 200 µm and 4 x 320 µm
-1x Fiber cutter
-50x Single-use tips including 1 x Bending tool
-2x Laser protective goggles for operator and assistant
-1x Laser protective goggles for patients
-1x Switching power supply
-1x Transport packaging